Revision A PRO 100-400 V2 Patient Monitor Service Manual 4-19
2009829-001
Calibration & Maintenance: SECTION 4. GENERAL MAINTENANCE
2. Power on the PRO Monitor and select More...
3. Select the Service mode and input 2 2 1 3.
4. From the Service menu, select the NIBP button.
5. From the NIBP menu, select the Check button.
6. Apply the following pressures (measured by an external digital
manometer) and confirm that the Monitor readings agree with the
following table for both PT1 and PT2 channels.
7. If calibration is required, refer to Section 4.7.1.
4.6.4.Neonate Overpressure Check
1. Set up the unit and calibration equipment as shown in Figure 4-1.
2. Power on the PRO Monitor and select More...
3. Select the Service mode and input 2 2 1 3.
4. From the Service menu, select the NIBP button, then select the
Popoff button.
Note: If the overpressure point is out of range, adjust the
overpressure potentiometer as described in 4.7.2.
5. Increase applied pressure until overpressure occurs. Confirm that
pressure at that point is between 150 mmHg to 165 mmHg and
system pressure falls to less than 20 mmHg within 4 seconds.
6. Power off the PRO Monitor.
7. If calibration is required, refer to Section 4.7.1.