Dinamap Pro 100V2 Wartungshandbuch Seite 49

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 104
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 48
Revision A PRO 100-400 V2 Patient Monitor Service Manual 3-9
2009829-001
Theory of Operation: 3.3 FUNCTIONAL DESCRIPTION
communications, and isolation between SPO2 PWA and Primary
Processor.
Patient data received from the finger sensor is filtered, amplified, and
analyzed on the SPO2 PWA. The information is sent to the Main
Board via the optically coupled electrically isolated serial connection.
The Primary Processor receives the data and distributes it to the
appropriate display. The data is also sent to the printer via the
Secondary Processor.
3.3.5 Pneumatic Control Device
The Main Board transducers PT1 and PT2 sense cuff pressure and
raw oscillatory information. The analog signals are amplified,
buffered, and sent to an Analog to Digital (A/D) Converter. The
converter changes the analog signals to digital signals and routes
these signals to the Primary Processor. The Primary Processor
receives the digital signals via the data interface.
Appropriate valve and pump control signals are sent from the Primary
Processor to the PCGAL. The PCGAL then generates the appropriate
drive signals to the pump and valves.
If a fail-safe mode or overpressure condition occurs, the PCGAL
provides the appropriate control signals to insure a safe condition,
where the cuff vents to atmosphere pressure.
3.3.6 LCD Assembly
The LCD assembly is used as a message center. It displays patient
vital signs, alarm status, monitor set-up, limit violation, BP cycle and
the time the data was received. The Primary Processor controls the
LCD assembly. When the parameter signals are read by the Primary
Processor, it decodes these signals and routes the digitized signals to
the graphic LCD.
The LCD assembly is equipped with a variable lighting feature for the
graphic display. The Primary Processor controls the lighting feature,
automatically. When a low or medium ambient lighting condition
occurs, the Primary Processor sends a BACKLIGHT signal to adjust
the graphic display lighting with respect to the room lighting condition.
If the lighting condition is high, the Primary Processor sends a
BACKLIGHT_1 signal to switch off the graphic display lighting.
3.3.7 Printer PWA w/Printer
The Printer PWA w/printer communicates to the Primary Processor
via Secondary Processor data line. The printer receives power from
the Main Board via the Printer PWA. The printer sends a DATA OUT
(CONTROL DATA) signal to the Secondary Processor to notify of it’s