Revision A PRO 100-400 V2 Patient Monitor Service Manual 2-9
2009829-001
Product Description: SECTION 2. PRODUCT DESCRIPTION
2.4. HOST PORT CONNECTOR (REAR PANEL)
WARNING! Auxiliary equipment connected to
the DINAMAP
®
PRO Series Monitor will result in
the formation of an electromedical system and
thus, must comply with the requirements of EN
60601-1-1/ IEC 601-1. All host port signals are
NON-ISOLATED and should be connected to
equipment conforming to IEC-601-1, configured
to comply with IEC 601-1-1 ONLY. Where isolation of data
communication is required, GE Medical Systems Information
Technologies part number ILC232 should be used. If external alarm
control is required, GE Medical Systems Information Technologies
part number 487208 (Isolated Remote Alarm Cable Assembly) should
ALWAYS be used. Please refer to the Information Sheet included
with the isolated remote alarm cable for details.
Note: When using remote alarm, the PRO Series Monitor should be
considered the primary alarm source. The secondary alarm is used
for secondary purposes only.
Pin Assignments
Pin Function
1 Ground
2 Inverted TTL Transmit Data
3 Inverted TTL Receive Data
4 Fused +5 Volts
5 No connection
6 No connection
7 Ground
8 Remote Alarm
9 RS232 Request to Send (RTS)
10 RS232 Clear to Send (CTS)
11 RS232 Transmit Data (TxD)
12 No connection
13 RS232 Receive Data (RxD)
14 No connection
15 No connection