2-14 PRO 100-400 V2 Patient Monitor Service Manual Revision A
2009829-001
Product Description: SECTION 2. PRODUCT DESCRIPTION
2.6.5. TURBO TEMP
Scale °Farenheit (F)
°Celsius (C)
Range
Predictive Mode Max: 41.1° C; 106.0° F
Min: 33.6° C; 96.0° F
Monitor Mode Max: 41.1° C; 106.0° F
Min: 26.7° C; 80.0°
Predictive Mode Accuracy ± 1.0° F
± 0.6° C
Monitor Mode Accuracy ± 0.1° C
± 0.2° F
(when tested in a calibrated liquid bath; meets ASTM E1112,
Table 1, in range specified)
Determination Time less than 60 seconds
Use only IVAC probes and probe covers. The size, shape, and thermal characteristics of the probe
covers can affect the performance of the instrument. Inaccurate readings or retention problems may
occur unless IVAC probes and probe covers are used.
IVAC
®
Patents
U.S. D300,728; D300,909
2.6.6. SpO
2
NELLCOR SpO
2
Measurement Range
SpO
2
1 to 100%
Pulse Rate 20 to 250 beats per minute
Perfusion Range 0.03 to 20%
Accuracy and Motion Tolerance
Saturation
Without Motion - Adults* 70 to 100% ±2 digits
Without Motion - Neonate* 70 to 100% ±3 digits
With Motion - Adults/Neo** 70 to 100% ±3 digits
Low Perfusion 70 to 100% ± 2 digits
Pulse Rate
Without Motion 20 to 250 beats/min ±3 digits
With Motion normal physiologic range
55 to125 beats/min ±digits