Revision A PRO 100-400 V2 Patient Monitor Service Manual 2-17
2009829-001
Product Description: SECTION 2. PRODUCT DESCRIPTION
>0.02% Pulse Amplitude
and % Transmission >5%
Saturation (% SpO2) ±2 digits
Pulse Rate ±3 digits
Interfering Substances
Carboxyhemoglobin may erroneously increase readings. The level of increase is approximately equal to the
amount of carboxyhemoglobin present. Dyes, or any substance containing dyes, that change usual arterial
pigmentation may cause erroneous readings.
Sensor Light Source
Wavelength Infrared: 905 nm (nominal)
Red: 660 nm (nominal)
Power Dissipation Infrared: 22.5 mW (max)
Red: 27.5 mW (max)
Default Settings
SpO2 (%) HIGH: 100
SpO2 (%) LOW: 90
Sensitivity Mode 2 (for low perfusion-Default)
Averaging Time 12 seconds
FastSAT Mode 0 (for Off)
Masimo Patents
5,482,036; 5,490,505; 5,632,272; 5,685,299; 5,758,644; 5,769,785; 6,002,952; 6,036,642; 6,067,462; 6,206,830; 6,157,850,
and international equivalents.