1DINAMAP® PROSeries 100-400V2MonitorOperation Manual

Seite 2 - PRO 400V2 Monitor

10Product ComplianceThe DINAMAP® PRO Monitor is classified in the following categories for compliance with IEC 601-1:• Class l, internally powered•Tra

Seite 3 - Operation Manual

100•Spray bottlesProcedure1. Prepare the enzymatic detergent according to the manufacturer’s instructions and the 10% bleach solution, in separate spr

Seite 4

101Appendix Fare damaged, they must be disposed of as advised in Appendix F.Storage and Battery CareIf it becomes necessary to store the Monitor for a

Seite 5 - Contents

102• Repeated failure to fully charge the battery will result in a significant reduction in battery life.• The expected lifetime of the battery largel

Seite 6

103Appendix FDisposal of Product Waste As you use the PRO Monitor, you will accumulate solid wastes that require proper disposal or recycling. These i

Seite 8

105Appendix GHost Port Connector (rear panel)WARNING! Auxiliary equipment connected to the DINAMAP® PRO Monitor will result in the formation of an ele

Seite 11 - Artifact

2009802-001 BEuropean RepresentativeGE Medical Systems Information Technologies GmbHPostfach 60 02 65D-79032 Freiburg, GermanyTel. +49 761 45 43 - 0Fa

Seite 12

11SymbolsThe following symbols are associated with the PRO Monitor. Note: The type of model determines which symbols appear on the Monitor. Attention,

Seite 13 - Getting Started

12Packaging label depicting the transportation and storage atmospheric pressure range of 500 to 1060 hPa.The DINAMAP® PRO Monitor is protected against

Seite 14

13Getting StartedUnpacking the Monitor and AccessoriesBefore attempting to use the PRO Monitor, take a few minutes to become acquainted with the Monit

Seite 15

14charging. If the battery is not inserted, the external power indicator LED will flash (short flash approx. every 4 sec). When the Monitor is running

Seite 16 - Rear Panel Connections

15Getting Startedin a medical/hospital grade attachment plug, provided with the following cord tag: “Hospital Grade Plug." Grounding integrity ca

Seite 17

16Rear Panel Connections1 Battery compartment cover: Retains and protects internal battery.2 Mains input: Used to connect to AC power supply.3 Externa

Seite 18

17Getting StartedFront Panel Controls and Indicators7 Systolic pressure display: 3-digit red LED indicates measured systolic BP in mmHg.8 Active temp

Seite 19

1814 External power indicator: Green LED indicates external power status and battery charging status of monitor.15 Temperature probe connector: Temper

Seite 20 - Using the Printer

19Getting StartedSwitching the Monitor On and OffTo switch the DINAMAP PRO Monitor on, press and hold the power ON/OFF switch (16) for at least 10 sec

Seite 21

2DINAMAP® PRO 400V2 Monitor

Seite 22 - Cautions

20appears dark on a light background, while the text of selected buttons appears light on a dark background.Note: Some menus have six option buttons.

Seite 23 - Using the Monitor

21Getting StartedAny time the paper is loaded, the printer automatically prints a test strip with the DINAMAP® PRO name on it. If no print is visible

Seite 24 - General Warnings

22StorageStore thermal paper in a cool, dry place. The printed strip (thermal paper recording) should not be• exposed to direct sunlight,• exposed to

Seite 25

23Using the MonitorNoninvasive Blood Pressure DeterminationDescriptionThe BP parameter is included in Models 100V2, 200V2, 300V2, and 400V2. Blood pre

Seite 26

24allowing SpO2 and predictive temperature determinations to be monitored and recorded (depending on Monitor model). In the Vitals (UK: All Obs) mode,

Seite 27

25Using the Monitorexamine the cuff site and the limb distal to the cuff regularly for signs of impeded blood flow.• Do not apply external pressure ag

Seite 28

26produce a peripheral pulse or the patient may have poor peripheral perfusion. Also, if a patient’s beat-to-beat pulse amplitude varies significantly

Seite 29

27Using the MonitorProcedures1. Connect the end of the air hose which has quick-release clips to the cuff connector (30) on the front of the Monitor.

Seite 30

285. Inspect cuff for damage. Replace cuff when aging, tearing, or weak closure is apparent. Do not inflate cuff when unwrapped.Precaution: Do not use

Seite 31

29Using the MonitorManual ModeTo start a determination, press the START/STOP BP key (29). A normal, uninterrupted Manual cycle takes about 40 seconds.

Seite 32

3DINAMAP® PRO MonitorOperation ManualThis manual is for DINAMAP® PRO Monitor Models 100V2, 200V2, 300V2, and 400V2, all with printers. • PRO 100V2: BP

Seite 33

30When Auto mode is selected, a number at the right of the Auto button indicates the time interval between each reading. To change the time interval,

Seite 34

31Using the Monitorinitial target pressure selection for neonates is 110 mmHg. Artifact rejection is relaxed in the Stat mode for adult/pediatric pati

Seite 35

32Temperature is shown on the temperature display in degrees Celsius or Fahrenheit, and the unit of measure is indicated by the °C °F display (13). Th

Seite 36

33Using the MonitorWhen a monitor mode measurement is in progress, it is terminated when the probe is inserted into the probe holster.Note: These temp

Seite 37

34readings may occur unless IVAC® probe covers are used. • If a patient’s temperature is below 96.0 °F, the unit will automatically switch from the n

Seite 38

35Using the Monitor7. Remove the probe. Discard the disposable cover by holding the probe as you would a syringe and pressing the button on the probe

Seite 39

36Procedures for Monitor Mode Determinations (Axillary Determinations)1. Connect the temperature probe cable to the temperature probe connector. 2. Re

Seite 40

37Using the Monitor• If the probe tip temperature is 92.0° F or higher (33.3° C) when taken out of the probe holster, the thermometer will not be able

Seite 41

38Pulse rate derived from SpO2 appears in the Pulse Rate window, and the SpO2 pulse indicator flashes synchronization with the real-time pulse rate me

Seite 42 - Warnings

39Using the MonitorIf the Monitor is unable to detect a pulse for 10 seconds during normal SpO2 monitoring, the values in the LED flash, alternating p

Seite 43

4other purposes or copying thereof is prohibited without prior written consent by GE Medical Systems Information Technologies, Tampa, Florida.Illustra

Seite 44

40- there is excessive illumination from light sources such as surgical lamp, a bilirubin lamp, or sunlight- a blood pressure cuff is inflated on the

Seite 45

41Using the Monitormeasurements. Always remove oximetry devices and attachments from the MRI environment before scanning a patient.• The use of cardio

Seite 46 - OxiMAX™ SpO

42• Use or patient-applied parts are latex-free.Procedures1. Select a sensor that is appropriate for the patient and the clinical situation.Warning: D

Seite 47

43Using the Monitor• Observe the sensor site to assure adequate distal circulation. Sensor sites should be checked at least every 2 hours and rotated

Seite 48 - General Cautions

44use a disposable adhesive sensor that may tolerate more motion.• Replace the sensor.PROBLEM: The SpO2 value or the pulse rate changes rapidly; the p

Seite 49

45Using the MonitorSOLUTION:• Check that calculations have been corrected appropriately for the relevant variable. In general, calculated saturation v

Seite 50

46• If there is excessive light, cover the sensor with opaque material.PROBLEM: A sensor error indicating a bad sensor has occurred.CAUSE:• The sensor

Seite 51

47Using the Monitorthat are derived from the SpO2 signal. A tone sounds at a rate corresponding to the SpO2 saturation level. The pitch is highest at

Seite 52

48normal SpO2 reporting of values when several consecutive good pulse determinations are made.General Warnings• Do not use the SpO2 function during ma

Seite 53

49Using the Monitor• Do not perform any testing or maintenance on a sensor while it is being used to monitor a patient.• Bright light sources (e.g., i

Seite 54

5Introduction ...7About the DINAMAP® PRO Monitor ...

Seite 55 - Using the Menu System

50pulsation and inaccurate measurement of oxygen saturation.• Excessive pressure from the sensor may cause necrosis of the skin.• For additional warni

Seite 56 - Normal Mode Menu

51Using the Monitor3. Plug the SpO2 sensor into the SpO2 sensor extension cable. Then plug the SpO2 sensor extension cable into the SpO2 sensor connec

Seite 57

52SOLUTION:Check the patient.• If possible, keep the patient still; check whether the sensor is applied securely and properly, and replace it if neces

Seite 58 - Main Menu

53Using the Monitor• If there is excessive light, cover the sensor with opaque material.• Circulation distal to the sensor site should be checked rout

Seite 59

54• The sensor or cable may be defective, or the cabling may be improperly connected.SOLUTION:Check the patient.• If possible, keep the patient still;

Seite 60

55Using the Menu SystemIntroductionThe PRO Monitor is equipped with a liquid crystal display (25) and a rotor (21). Used together, these allow the ope

Seite 61

56Normal Mode Menu

Seite 62

57Using the Menu SystemClinician Mode Menu

Seite 63

58RotorRotating the rotor causes option buttons to be highlighted (light text on a dark background). Turning the rotor produces a click. Turning it cl

Seite 64

59Using the Menu SystemVitals Button (UK: All Obs)Selection of this button initiates a BP determination while allowing SpO2 and predictive temperature

Seite 65

6Appendix A...77Technical Specifications...

Seite 66 - More... Menu

60Note: If the SpO2 plethysmograph is displayed on the LCD, the waveform pauses for 2 minutes or until the Clear button is selected. SpO2 values are a

Seite 67

61Using the Menu Systemoperating mode returns to the default setting of Manual. The default setting of Manual can be overridden to return to the previ

Seite 68

62Alarms ButtonSelection of this button displays the Alarms menu. This menu is used to adjust the violation limits for BP, Pulse Rate, and SpO2. The v

Seite 69

63Using the Menu SystemAutoSelection of this button updates the alarm limits relative to the current parameter values. Pressing this button cancels an

Seite 70

64DisplaySelection of this button allows the operator to view the trend data. Note: If the trend data have been lost (e.g., if the clock settings have

Seite 71

65Using the Menu SystemMain Selection of this button returns the user to the Main menu.Print ButtonSelection of this button displays the Print menu.Au

Seite 72

66More... MenuThis menu is used to set the various operating modes of the Monitor.SpO2 Button (Models 300V2 and 400V2)Selection of this button display

Seite 73

67Using the Menu SystemResponse Mode (Nellcor Only)Response mode allows the user to specify the averaging technique to optimize measurements in the pr

Seite 74

68Pwr Sav (Sleep Mode)Selection of this button allows the operator to specify the time, in minutes, that elapses before the Monitor goes into “sleep”

Seite 75

69Using the Menu SystemTimeSelection of this button allows the operator to change the internal time and date of the Monitor. The clock, which is maint

Seite 76 - Main menu

7About the DINAMAP® PRO MonitorDINAMAP® PRO Monitors provide noninvasive determination of systolic blood pressure, diastolic blood pressure, mean arte

Seite 77 - Appendix A

70Main Selection of this button returns the user to the Main menu.TempSelection of this button displays the temperature submenu, which allows the user

Seite 78

71Using the Menu SystemSpO2 PlethWhen this option is checked and SpO2 data are available, the plethysmograph waveform and the pulse amplitude bar will

Seite 79

72Clinician MenuPress. Selection of this button displays a panel for setting the default BP target inflation pressure. Adjusting the default target pr

Seite 80

73Using the Menu SystemOK. Selection of this button returns the user to the Clinician menu.More... Selection of this button displays the More... menu

Seite 81

74For monitors with Nellcor SpO2 technology, the SpO2 configuration menu includes Response Mode and SatSeconds™. Response Mode can be set to “1” or “

Seite 82

75Using the Menu SystemSilence. Selection of this button will cause all alarms except the FAILSAFE alarm to be muted. A confirmation menu will appear

Seite 83

76OK ButtonSelection of this button acknowledges the error. The Monitor clears the identified error and then returns the user to the Main menu.

Seite 84

77Appendix ATechnical SpecificationsBPCuff Pressure Range 0 to 290 mmHg (adult)(Normal operating range) 0 to 140 mmHg (neonate)Default Target: Cuff In

Seite 85 - Appendix B

78TURBOTEMP TemperatureScale °Fahrenheit (F)°Celsius (C)RangePredictive mode Max: 41.1° C; 106.0° FMin: 33.6° C; 96.0° F Monitor mode Max: 41.1° C;

Seite 86 - Hierarchy of Alarms

79Appendix APulse RateWithout Motion 20 to 250 beats/min ±3 digits With Motion normal physiologic range 55 to 125 beats/min ±5 digits Low Perfusion 20

Seite 87

8been reports of sensors causing patient burns when operating in an MRI environment.• Do not use the Monitor in the presence of flammable anesthetics.

Seite 88

80Nellcor Patents4,621,643; 4,653,498; 4,700,708; 4,770,179; 4,802,486; 4,869,254; 4,928,692; 4,934,372; 5,078,136; 5,351,685; 5,421,329; 5,485,847; 5

Seite 89

81Appendix Astandard deviation. Plus or minus one standard deviation encompasses 68% of the population.‡The Masimo SET® SpO2 parameter has been valida

Seite 90

82Default Settings SpO2 (%) HIGH: 100 SpO2 (%) LOW: 90Sensitivity Mode 2 (for low perfusion-Default)Averaging Time 12 secondsFastSAT Mode 0 (for Off)M

Seite 91 - Appendix C

83Appendix Asupplied from a source conforming to IEC 601-1.AC input is protected by two internal fuses, replaceable by qualified service personnel onl

Seite 92

84Compatibility Requirements and Tests and CISPR 11 (Group 1, Class A) for radiated and conducted emissions.IPX1The DINAMAP® PRO Monitor is protected

Seite 93

85Appendix BAlarm CodesAll alarm indications are accompanied by an audible signal unless Alarm Silence is selected.A microprocessor system failure wil

Seite 94

86Hierarchy of AlarmsAlarms in the DINAMAP® PRO Monitor are in three priority levels. They are:AlarmPriority LevelFailsafe 1Patient and system 2 (High

Seite 95 - Appendix D

87Appendix BProcedural and Error Alarm CodesAlarm CodeLED DisplayLCD DescriptionAudible Tone and VolumeEffect of Alarm Silence SwitchEffect of Clear v

Seite 96

88Procedural and Error Alarm Codes (cont.)Alarm CodeLED DisplayLCD DescriptionAudible Tone and VolumeEffect of Alarm Silence SwitchEffect of Clear via

Seite 97 - Appendix E

89Appendix BProcedural and Error Alarm CodesAlarm CodeLED DisplayLCD DescriptionAudible Tone and VolumeEffect of Alarm Silence SwitchEffect of Clear v

Seite 98 - Service Manuals

9presence of equipment which does not conform to these specifications.• Place the PRO Monitor on a rigid, secure surface. Monitor must only be used wi

Seite 100

91Appendix CPrinciples of Noninvasive Blood Pressure DeterminationThe oscillometric method of determining NIBP is accomplished by a sensitive transduc

Seite 101 - Appendix F

92Full NIBP Determination Sequence for AdultAt each step the microprocessor stores cuff pressure, the matched pulse amplitude, and the time between su

Seite 102 - Leak Testing

93Appendix CNIBP - Auto ModeSystolic SearchIf systolic pressure is not found, the NIBP parameter can search at cuff pressures higher than the initial

Seite 103

94make an accelerated determination of blood pressure if it has been 16 minutes or less since the last determination.

Seite 104

95Appendix DReorder CodesPROD PRODUCTCODE ECATDINAMAP PRO Monitor Operations Manual-English 2009802-001DINAMAP PRO Monitor Service Manual 2009829-001B

Seite 106

97Appendix EWarranty, Service, and Spare PartsWarning: There are no user serviceable parts inside the DINAMAP® PRO Monitor. Refer all servicing to qua

Seite 107

98RepairsIf your product requires warranty, extended warranty, or non-warranty repair service, call GE Medical Systems Information Technologies and a

Seite 108

99Appendix FMaintenanceCleaning the MonitorThe Monitor and accessories are to be kept clean and used according to the instructions provided here and i