80
Nellcor Patents
4,621,643; 4,653,498; 4,700,708; 4,770,179; 4,802,486;
4,869,254; 4,928,692; 4,934,372; 5,078,136; 5,351,685;
5,421,329; 5,485,847; 5,533,507; 5,577,500; 5,803,910;
5,853,364; 5,865,736; 6,083,172; Re. 35,122 and,
international equivalents.
Masimo SET SpO
2
Measurement Range
SpO
2
1 to 100%
Pulse Rate 25 to 240 beats/min
Perfusion Range 0.02 to 20%
Accuracy and Motion Tolerance
Saturation
Without Motion - Adult/Ped* 70 to 100% ±2 digits
Without Motion - Neonate* 70 to 100% ±3 digits
With Motion - Adult/Ped/Neo**† 70 to 100% ±3 digits
Low Perfusion‡ 70 to 100% ±2 digits
0 to 69% unspecified
Pulse Rate
Without Motion 25 to 240 beats/min ±3 digits
With Motion normal physiologic range
25 to 240 beats/min ±5 digits
*The Masimo SET
®
SpO
2
parameter with LNOP-Adt sensors has been
validated for no motion accuracy in human blood studies on healthy adult
volunteers in induced hypoxia studies in the range of 70-100% SpO
2
against
a laboratory co-oximeter and ECG monitor. This variation equals plus or
minus one standard deviation. Plus or minus one standard deviation
encompasses 68% of the population.
**The Masimo SET
®
SpO
2
parameter with LNOP-Adt sensors has been
validated for motion accuracy in human blood studies on healthy adult
volunteers in induced hypoxia studies while performing rubbing and tapping
motions at 2 to 4 Hz at an amplitude of 1 to 2 cm and a non repetitive
motion before 1 to 5 Hz at an amplitude of 2 to 3 cm in induced hypoxia
studies in the range of 70-100% SpO2 against a laboratory co-oximeter and
ECG monitor. This variation equals plus or minus one standard deviation.
Plus or minus one standard deviation encompasses 68% of the population.
†The Masimo SET
®
SpO
2
parameter with LNOP-Neo Pt sensors has been
validated for neonatal motion accuracy in human blood studies on neonates
while moving the neonate’s foot at 2 to 4 cm against a laboratory co-
oximeter and ECG monitor. This validation equals plus or minus, one