79
Appendix A
Pulse Rate
Without Motion 20 to 250 beats/min ±3 digits
With Motion normal physiologic range
55 to 125 beats/min ±5 digits
Low Perfusion 20 to 250 beats/min ±3 digits
*Adult specifications are shown for OXIMAX MAX-A and MAX-N sensors.
Neonate specifications are shown for OXIMAX MAX-N. Saturation accuracy
will vary by sensor type.
**Applicability: OXIMAX MAX-A, MAX-AL, MAX-P, MAX-I, and MAX-N
sensors.
Default Settings
SpO
2
(%) HIGH: 100
SpO
2
(%) LOW: 90
Response Mode 2 (for Mode 2: Fast Response)
Sat Seconds 0
Audible Indicator Pitch changes continuously
with saturation; volume from 0
(off) to 9
Waveforms Pulse plethysmograph
waveform on LCD gain
compensated
Sensor Connect/ Monitor detect attachment
Disconnect From Patient or disconnection of sensor
from patient within 15 s
Pulse Detection Monitor will detect pulse or
enter no signal state within 15
s of being attached to patient
Loss of Pulse Monitor will detect loss of
pulse from patient and enter
no signal state within 10 s
Sensor Light Source
Wavelength Infrared: 890 nm (nominal)
Red: 660 nm (nominal)
Power Dissipation 52.5 mW (max)